<ErrorNumber>10061</ErrorNumber>
<ErrorFamily>Hardware</ErrorFamily>
<ErrorMessage>
HARDWARE: External Debris Blower Filters Full
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The debris cabinet is reporting that the filters are full.  Check the filters, and replace as
necessary.
</text></DescriptionText>
<DescriptionText><text>
If the filter looks fine, check the sensor inside the cabinet, and the cable running to the device.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
The device thinks the debris filter inside the debris cabinet is full.  Open the cover of the
cabinet and replace the filter.  If the problem comes back, contact the CreoScitex Response Centre.
</text></DescriptionText>
</OperatorAction>
<Cause>
Filter is full.
</Cause>
<Cause>
Faulty cable from debris cabinet to device.
</Cause>
<Cause>
Faulty sensor inside debris cabinet.
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/dr.xml</LinkAddress><LinkTitle>Command: dr</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:32:31 2008 PST
</Updated>
<Links>
</Links>
</Error>